Forty-one studies were included in the review (n=2,515; range 20 to 133). Of these, 27 reported consecutive enrollment and 34 blinding to angiography findings when interpreting CE-CCT.
For all CT
The pooled sensitivity was 95 per cent, specificity was 85 per cent and DOR 102 (95% CI: 63, 141) for all segments. The pooled sensitivity was 96 per cent, specificity 95 per cent and DOR 429 (95% CI: 240, 617) for assessable segments. The pooled sensitivity was 99 per cent, specificity 76 per cent and DOR 306 (95% CI: 183, 428) per patient.
For electron beam CT
The pooled sensitivity was 80 per cent, specificity 77 per cent and DOR 13 (95% CI: -52, 77) for all segments. The pooled sensitivity was 86 per cent, specificity 90 per cent and DOR 60 (95% CI: -282, 401) for assessable segments. The pooled sensitivity was 90 per cent, specificity 59 per cent and DOR 12 (95% CI: -228, 252) per patient.
For multidetector CT
The pooled sensitivity was 95 per cent, specificity 90 per cent and DOR 152 (95% CI: 103, 200) for all segments. The pooled sensitivity was 96 per cent, specificity 96 per cent and DOR 591 (95% CI: 365, 817) for assessable segments. The pooled sensitivity was 99 per cent, specificity 85 per cent and DOR 408 (95% CI: 266, 550) per patient.
Results for different types of multidetector CT and sensitivity analyses were presented.